10.3389/fneur.2018.00558.s002 Marnix Naber Marnix Naber Carlien Roelofzen Carlien Roelofzen Alessio Fracasso Alessio Fracasso Douwe P. Bergsma Douwe P. Bergsma Mies van Genderen Mies van Genderen Giorgio L. Porro Giorgio L. Porro Serge O. Dumoulin Serge O. Dumoulin Image_2_Gaze-Contingent Flicker Pupil Perimetry Detects Scotomas in Patients With Cerebral Visual Impairments or Glaucoma.PDF Frontiers 2018 pupillary response perimetry open-angle glaucoma cerebral visual impairment pupillometry neuro-ophthalmic disease visual field defect 2018-07-10 04:26:07 Figure https://frontiersin.figshare.com/articles/figure/Image_2_Gaze-Contingent_Flicker_Pupil_Perimetry_Detects_Scotomas_in_Patients_With_Cerebral_Visual_Impairments_or_Glaucoma_PDF/6794858 <p>Background: The pupillary light reflex is weaker for stimuli presented inside as compared to outside absolute scotomas. Pupillograph perimetry could thus be an objective measure of impaired visual processing. However, the diagnostic accuracy in detecting scotomas has remained unclear. We quantitatively investigated the accuracy of a novel form of pupil perimetry.</p><p>Methods: The new perimetry method, termed gaze-contingent flicker pupil perimetry, consists of the repetitive on, and off flickering of a bright disk (2 hz; 320 cd/m<sup>2</sup>; 4° diameter) on a gray background (160 cd/m<sup>2</sup>) for 4 seconds per stimulus location. The disk evokes continuous pupil oscillations at the same rate as its flicker frequency, and the oscillatory power of the pupil reflects visual sensitivity. We monocularly presented the disk at a total of 80 locations in the central visual field (max. 15°). The location of the flickering disk moved along with gaze to reduce confounds of eye movements (gaze-contingent paradigm). The test lasted ~5 min per eye and was performed on 7 patients with cerebral visual impairment (CVI), 8 patients with primary open angle glaucoma (age >45), and 14 healthy, age/gender-matched controls.</p><p>Results: For all patients, pupil oscillation power (FFT based response amplitude to flicker) was significantly weaker when the flickering disk was presented in the impaired as compared to the intact visual field (CVI: 12%, AUC = 0.73; glaucoma: 9%, AUC = 0.63). Differences in power values between impaired and intact visual fields of patients were larger than differences in power values at corresponding locations in the visual fields of the healthy control group (CVI: AUC = 0.95; glaucoma: AUC = 0.87). Pupil sensitivity maps highlighted large field scotomas and indicated the type of visual field defect (VFD) as initially diagnosed with standard automated perimetry (SAP) fairly accurately in CVI patients but less accurately in glaucoma patients.</p><p>Conclusions: We provide the first quantitative and objective evidence of flicker pupil perimetry's potential in detecting CVI-and glaucoma-induced VFDs. Gaze-contingent flicker pupil perimetry is a useful form of objective perimetry and results suggest it can be used to assess large VFDs with young CVI patients whom are unable to perform SAP.</p>